Ext.
Select Yes if you connect an external humidifier.
humidifier
View
Displayed at all times when an oximeter is connected.
oximeter
Options
Parameter
Description
Essentials
Set the level of access available to the patient.
Leak Alert
Enable / disable the Leak Alert feature. When enabled, leaks >40 L/min
(0.7 L/s) for >20 sec result in an audible alert and a high leak message is
displayed.
™
SmartStart
Enable / disable the SmartStart feature. If you enable the SmartStart
feature, the device will start automatically when the patient breathes into
the mask and then stop automatically when the patient removes the
mask.
Reminders
Mask
Set a recurring reminder to the patient to replace the mask.
Humidifier
Set a recurring reminder to the patient to replace the humidifier.
Tube
Set a recurring reminder to the patient to replace the air tubing.
Filter
Set a recurring reminder to the patient to replace the air filter.
Configuration
Parameter
Description
Language
Set the display language.
(Not all languages available in all regions.)
Date
Set the current date.
If you set a new date that occurs in the past then an error message is
displayed. Before this change can be made, erase the compliance data
available under the Configuration menu.
Time
Set the current time.
If you set a new time that occurs in the past then an error message is
displayed. Before this change can be made, erase the compliance data
available under the Configuration menu.
12
No / Yes
0-300 hrs
0-100% SpO
2
Range
On / Plus
Off / On
Off / On
Off / 1– 24 mths, 1 month
increments
Off / 1–24 mths, 1 month
increments
Off / 1–24 mths, 1 month
increments
Off / 1–24 mths, 1 month
increments
Selection
English / Français / Español /
Português / Deutsch /
Italiano / Nederlands /
Svenska / Norsk / Dansk /
Suomi / Polski / Türkçe /
Русский / 简体中文 /
繁體中文
DD Mmm YYYY
24 hours