Sleep Report Screen Parameters - ResMed Defy Mini User Manual

Clinical guide
Table of Contents

Advertisement

 

Sleep Report screen parameters

Parameter
Usage hours
Events (AHI) per hour
Mask Seal
Humidifier
More Info
Period
Days Used
Days 4hrs+
Avg. Usage
Used Hrs
Pressure
Exp. Pressure
Leak
AHI
Total AI
AHI and Total AI: Central sleep apnoea detection (CSAD) is not active in ASVAuto and ASV modes. The ASV algorithm
eliminates central apnoeas. Therefore, any apnoeas reported by the device will be obstructive or indicative of a closed
airway.
Description
Number of hours the device has been used during the last session.
Apnoeas and hypopnoeas measured per hour for one day. An apnoea is when the
respiratory flow decreases by more than 75% for at least 10 sec. A hypopnoea is when
the respiratory flow decreases to 50% for at least 10 sec. The Apnoea Index (AI) and
Apnoea-Hypopnoea Index (AHI) are calculated by dividing the total number of events
that occurred by the total mask-on therapy period in hours.
th
Good—if the 70
percentile leak is less than 24 L/min.
Mask needs adjustment.
Humidifier attached and functional.
Humidifier fault; refer to troubleshooting section.
Set the time interval covered by the Sleep Report.
The options are: 1 Day / 1 Week / 1 Month / 3 Months / 6 Months / 1 Year
Number of days the device has been used during the selected period or since the last
compliance data was reset.
Number of days the device has been used for more than 4 hours during the selected
period or since the last compliance data was reset.
Average number of hours per day the device has been used during the selected period.
Number of hours the device has been used during the selected period or since the last
compliance data reset.
Average inspiratory pressure during the selected period (95
th
average of 95
percentile values for periods >1 day).
Average expiratory pressure during the selected period (95
th
average of 95
percentile values for periods >1 day).
th
Average of the 95
percentile values of leak during the selected period for days with
usage only.
Apnoea-Hypopnoea Index—average AHI during the selected period. AHI and AI are
calculated for times of low leak only.
Apnoea Index—average total AI during the selected period.
th
percentile for each day;
th
percentile for each day;