Viewing the Sleep Report

The Sleep Report screen shows sleep quality and mask seal status for the most recent therapy
session. Turn the dial to scroll down to view more detailed usage data. The parameters displayed
will depend on the therapy mode.

Sleep Report screen parameters

Parameter
Usage hours
Events (AHI) per hour
Mask Seal
Humidifier
More Info
Period
Days Used
Days 4hrs+
Avg. Usage
Used Hrs
28
Description
Number of hours the device has been used during the last session.
Apneas and hypopneas measured per hour for one day. An apnea is when the
respiratory flow decreases by more than 75% for at least 10 sec. A hypopnea is when
the respiratory flow decreases to 50% for at least 10 sec. The Apnea Index (AI) and
Apnea-Hypopnea Index (AHI) are calculated by dividing the total number of events that
occurred by the total mask-on therapy period in hours.
th
Good—if the 70
percentile leak is less than 24 L/min.
Mask needs adjustment.
Humidifier attached and functional.
Humidifier fault; refer to troubleshooting section.
Set the time interval covered by the Sleep Report.
The options are: 1 Day / 1 Week / 1 Month / 3 Months / 6 Months / 1 Year
Number of days the device has been used during the selected period or since the last
compliance data was reset.
Number of days the device has been used for more than 4 hours during the selected
period or since the last compliance data was reset.
Average number of hours per day the device has been used during the selected period.
Number of hours the device has been used during the selected period or since the last
compliance data reset.