SECTION 7—TROUBLESHOOTING
PROBLEM
No indicator lights on when
compressor turned on.
RED Light ON
Compressor cycles ON and Off
or
O
BELOW NORMAL
2
(YELLOW) and FILLING
(GREEN) lights are alternating
or
O
BELOW NORMAL
2
(YELLOW) light goes On and Off
Part No 1100873
SECTION 7—TROUBLESHOOTING
TROUBLESHOOTING
1. Turn compressor Off.
2. Check that power cord is properly plugged into
compressor and wall outlet
3. Turn the compressor On.
4. Clean air filters. Refer to Cleaning the Compressor
Filter on page 33.
5. If performing STEPS 1-4 does not cause any indicator
lights to come on, contact your home health care pro-
vider or Invacare for service.
1.Check the cylinder gauge to see if it is FULL:
• Cylinder full - Remove the cylinder and turn the unit
Off. Perform the
Cylinder Prefill Inspection on page 20 with a cylinder
that is NOT FULL.
• Cylinder not full - Proceed to STEP 2.
2. Check that the cylinder is connected properly.
3. Turn the compressor Off.
4. Remove the cylinder and reinstall.
5. Turn the compressor On. If cylinder is not full and is
connected properly, the compressor may have internal
failure. Contact your home health care provider or
Invacare.
1. Check the concentrator liter flow
setting is at 3 L/min or less.
If necessary, adjust the liter flow setting (refer to the
concentrator owner's manual). Wait at least 20 min-
utes for the GREEN light to appear. If it does not
appear, proceed to STEP 2.
2. The concentrator may need service, contact your
home health care provider or Invacare.
31
SOLUTION
.
HomeFill
Compressor
®