Output Control Keys - Agilent Technologies 6631B User Manual

Dynamic measurement dc source/system dc power supply
Hide thumbs Also See for 6631B:
Table of Contents

Advertisement

5 - Front Panel Operation

Output Control Keys

Output control keys control the output functions of the dc source.
Voltage
Press this key to access the voltage menu.
Display
VOLT <value>
Current
Press this key to access the current menu.
Display
CURR <value>
Output
Press this key to access the output menu list.
Display
*RST
PON:STATE <char>
RI <char>
DFI <char>
DFI:SOUR <char>
PORT <char>
DIGIO <char>
RELAY <char>
REL:POL <char>
PROT:DLY <value>
Protect
Press this key to display protection status.
Display
OC OT OV RI FS
-- -- -- -- --
Shift
OV
Press this key to access the overvoltage protection menu.
Display
VOLT:PROT <value>
This key accesses the calibration menu (Refer to Appendix B to calibrate your
Shift
Cal
dc source).
Notes:
1
These parameters are stored in non-volatile memory
2
These status summary bits are explained in chapter 3 of the Programming Guide
value = a numeric value
char = a character string parameter
p
Use
Ë
Use
É
Use
40
Command Function
Sets the output voltage
Command Function
Sets the output current
Command Function
Places the dc source in the factory-default state
Select the power-on state command (RST or RCL0)
Sets the remote inhibit mode (LATCHING, LIVE, or OFF)
Sets the discrete fault indicator state (ON or OFF)
Selects the DFI source (QUES, OPER, ESB, RQS, or OFF)
Sets the output port functions (RIDFI or DIGIO)
Sets and reads the I/O port value (0 through 7)
Set relay state independent of output (''ON'' or ''OFF'')
Set relay polarity (''NORM'' or ''REV'').
Sets the output protection delay in seconds
Command Function
Status of the protection features (example shows all tripped)
Status of the protection features (example shows none tripped)
Command Function
Sets overvoltage protection level
q
and
to scroll through the menu commands.
Ì
and
to scroll through the menu parameters.
Ê
and
to select a digit in a numeric entry field.
1
1
2
1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6633b66332a6634b6632b

Table of Contents