Replacing the Mast Actuator

DO NOT overtighten the nut and bolt. This damages the mounting bracket.
NOTE: For this procedure, refer to FIGURE 7.6.
1. Remove the bottom nut, washer and shoulder bolt that secure the mast actuator to the 
mast mounting bracket.
2. Rest the lift arm on your shoulder and remove the top nut, bolt, bracket and bushing 
from the lift arm mounting bracket.
3. Remove the mast actuator.
4. Reverse the above steps for installation of the new mast actuator.
Lift Arm Mounting Bracket
Shoulder Bolt
Part No. 1145811
CAUTION
Top Nut
FIGURE 7.6 Replacing the Mast Actuator
35
Bracket
Top Bolt
Bushing
Mast Actuator
Bottom Nut
Washer
Mast Mounting Bracket