Symbols - A&D UA-705 Instruction Manual

Hide thumbs Also See for UA-705:
Table of Contents

Advertisement

Symbols

Standby and ON switch.
Battery installation guide
SN
Serial number
Date of manufacture
Type BF: Device, cuff and tubing are
designed to provide special protection
against electrical shocks.
Indicates measurement is in progress.
It blinks while detecting a pulse.
Irregular Heartbeat indicator. (I.H.B.)
Indicator shows when an irregular
heartbeat or excessive body
movement is detected during the
measurement.
Flashes when there is air remaining in
the cuff at power up.
Flashes at the completion of the
measurement until exhaust is complete.
Flashes when pressure is insufficient
for the measurement.
Previous measurements stored in
memory.
Indicates remaining battery power.
Full Battery
Blinks when battery is low.
low Battery
Unstable blood pressure due to
movement during the measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during inflation.
The cuff is not fastened correctly.
The pulse is not detected correctly.
PUL. DISPLAY ERROR
EC directive medical device label
Symbols
Function / Meaning
English 4
Recommended Action
Measurement is in
progress. Remain as still
as possible.
Press
the
exhaust
valve button to exhaust
air from the cuff.
Pump air into the cuff
with the rubber bulb.
Replace battery with
new one when the
indicator blinks.
Try measurement again.
Remain very still during
the measurement.
Check the air hose
connecter plug and cuff.
Fasten the cuff correctly,
and try the measurement
again.