A&D UA-767 Instruction Manual

Digital blood pressure monitor
Hide thumbs Also See for UA-767:

Advertisement

Digital Blood Pressure Monitor
Model UA-767
Instruction Manual
Manuel d'instructions
Manual de Instrucciones
Manuale di Istruzioni
U4+5659G

Advertisement

Table of Contents
loading

Summary of Contents for A&D UA-767

  • Page 1 Digital Blood Pressure Monitor Model UA-767 Instruction Manual Manuel d’instructions Manual de Instrucciones Manuale di Istruzioni U4+5659G...
  • Page 2: Dear Customers

    ENGLISH Dear Customers Congratulations. You have purchased an A&D blood pressure monitor, one of the most technologically advanced, yet easy to use products available in the marketplace today. We strongly recommend you read this instruction manual carefully prior to using it the first time. General Information This device conforms to the European Directive 93/42/EEC for Medical Products.
  • Page 3: Parts Identification

    Parts Identification ENGLISH 2...
  • Page 4 Symbols Symbols Function / Meaning Treatment Turns the device on and off. ___________ Guide to install battery ___________ Direct current ___________ Serial number ___________ Date of manufacture ___________ Type BF: Device, cuff and tubing are designed to provide special ___________ protection against electrical...
  • Page 5: Using The Monitor

    Using the Monitor 1. Installing / changing the batteries Remove the battery cover and insert new batteries into the battery compartment as shown, taking care that the polarities (+) and (-) are observed. Use only R6P, LR6, AA or similar type batteries. •...
  • Page 6: Attaching The Arm Cuff

    Using the Monitor 3. Attaching the arm cuff Wrap the cuff around the upper arm, about 2-3 cm above the elbow, as shown. Place the cuff directly against the skin, as clothing may cause a faint pulse, and result in a measurement error.
  • Page 7 Measurements 1. Place the cuff on the arm (preferably the left arm). 2. Set the PRESSURE switch to a value that will be 30 to 40 mmHg above the expected systolic value and press the START button. • When the START button is pressed, all of the display symbols will appear for about one second.
  • Page 8 Measurements • Remain still while the device is in operation. • When a measurement is made with insufficient pressure, the device will pump again to a higher pressure. 4. When the measurement is complete, the buzzer sounds and the air is automatically released from the cuff.
  • Page 9: Notes For Proper Measurement

    Measurements • Accurate measurement cannot be made if the exhaust velocity is outside the range of 2 to 5 mmHg. This can be determined by watching the display each time the "Measurement in Progress" mark flashes. Note the reading as the "Measurement in Progress" mark flashes. Notes for proper measurement Sit in a comfortable position.
  • Page 10: About Blood Pressure

    About Blood Pressure 1. What is blood pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg). One's natural blood pressure is represented by the fundamental pressure, which is measured first thing in the morning while one is still at rest and before taking food or drink.
  • Page 11: Who Blood Pressure Classification

    About Blood Pressure 4. WHO blood pressure classification Standards for assessment of high blood pressure, without regard to age, have been established by the World Health Organization (WHO), as shown in the chart below. 5. Blood pressure variations An individual’s blood pressure varies greatly on a daily and seasonal basis. It may vary by 30 to 50 mmHg due to various conditions during the day.
  • Page 12: Troubleshooting

    Troubleshooting Problem Probable cause Corrective action Nothing appears Batteries are drained. Replace all batteries with new in the display, ones. even when the Battery terminals are not in Place batteries with power is turned the correct position. negative and positive terminals matching those indicated in the battery compartment.
  • Page 13: Maintenance

    Technical Data Type UA-767 Measurement method Oscillometric measurement Measurement range Pressure : 20 - 280 mmHg...