Deleting Old Data

Caution
Confirm that the data has already been transferred and saved, when
the data is to be deleted. It is not possible to recover data that was
deleted.
It is not possible to delete data completely, if the
released while the buzzer sounds at Step 2.
Steps for deleting old data
Press and hold the
Step 1
If you want to cancel this process, press the
Press and hold the
Step 2

Resetting the Recorder

If the recorder does not work correctly, press the reset key. The recorder
deletes all data and parameters. The internal system is initialized.
Caution
All data and parameters are deleted and preset initial parameters by
the reset.
Do not press the reset switch to intensely. Press this key gently so
as not to damage the components inside.
Keep foreign matter away from the reset switch hole.
Steps for reset
Open the battery cover.
Step 1
Turn off the power switch.
Step 2
Remove the batteries from the recorder.
Step 3
Press the reset key gently.
Step 4
Place new batteries in the recorder.
Step 5
Turn on the power switch. The recorder sounds the
Step 6
buzzer four times and
Set the parameters for the display and clock. Also,
Step 7
adjust the parameters for automatic measurement.
key for approx. 9 seconds.
START STOP
key once more until the buzzer becomes silent.
START STOP
is displayed blinking.
Page 22
START STOP
key.
AUTO ON/OFF
Reset key
key is
is displayed.