When a "B" is displayed, the backup batteries that is preserving a patient's data
are weak. Please transfer the data and save it to other media as soon as possible.
ID number
If you reset the recorder, the ID number is set to "1".
The ID number can be set using the optional software.
Performance specifications
Measurement method
Pressurization
Measurement range
Accuracy
Minimum display division
Depressurization
Measurement
Number of measurement Apporx. 200 times (It may reduce this number due to
Memory
Display
Clock
Batteries
Type of protecting
against electric shock
CE Marking
C-Tick Marking
AAMI : Association for the Advancement of Medical Instrumentation
ACA: the Australian Communications Authority
CE marking and C-Tick marking are labeled only where they are required.
Oscillometric
micro-pump
Display range
Automatic measurement
Manual measurement
Systolic Blood Pressure
Diastolic Blood Pressure
Pulse rate
Pressure
Blood pressure
Pulse rate
Pressure
Pulse rate
Constant exhaust
Exhaust
Automatic measurement
Manual measurement
environment and capacity of Ni-Cd batteries)
Up to 300 sets of data
Normal
During a measurement
After a measurement
Error code, function of concealing the measurement data
24-hours (1997~2096year, automatic leap year setting)
3 x Alkaline battery (type LR6, type AA, Mignon) or
3 x Ni-Cd battery (type AA, Mignon)
Internally powered equipment
The label of the medical device
by the EC directive.
The certification trade mark registered
to the ACA by the Trademark office.
Page 11
0 ~ 320 mmHg
85 ~ 300 mmHg(Fitted)
185 mmHg (Fixed)
60 ~ 280 mmHg
40 ~ 60 mmHg
30 ~ 200 bpm
+ 3 mmHg
Conforming to 1992 AAMI standard
+ 5 %
1 mmHg
1 bpm
Controlled ceramic valve
Ceramic valve
Clock
Pressure value
SYS, DIA and pulse rate
type BF
0366
N92