The following example demonstrates the rules for setting up a time-gated measurement. In this
example, we are using two signal generators to generate two signals at the same frequency
(50 MHz). The pulse generators "space" (interleave) the signals in time as well as pulse
modulate the signals.
E Y T
R IG OUTPUl
PULSE
GENERATOR #2
Note
Be sure that the input impedance for the oscilloscope channels is set to 1 MO.
Setting
Period
Width
Trigger
Voltage
(peak-to-peak)
Trigger delay
M O D U L A T I O N
SIGNAL
GENERATOR #2
P U L S E
RF
M O D U L A T I O N
O U T P U T
R E F L E C T E D
SOURCE
D I R E C T I O N A L
BRIDGE
Figure 4-21. Test Setup for Option 105
Pulse
Generator #l
280 ps
50 /a
Positive edge of square wave
5 v
85 ps
OSCILLOSCOPE
CH
3
CH
2
H 4
GATE
N P U T
Pulse
Generator #2
280 /Is
50 us
Not applicable
5 v
None
Making Measurements 4-27