1.0 Overview of the 8800 Series Trace Moisture Analyzer

The 8800 Series Trace Moisture Analyzer is a microprocessor based hygrometer, for measuring
moisture content in gases in the range from -100°C to +20°C dewpoint depending on the sensor
ordered. The 8800 Series uses a Hyper Thin Film HTF™ sensor which is encapsulated in sintered
stainless steel, thus it is capable of coming into contact with a wide variety of environments.
However one should keep in mind that the sensor is a delicate device and it should be handled
accordingly.
The measurement is displayed on the instrument's custom LCD, can be transmitted by optional
analog and digital outputs, and can control optional programmable relays. Four front panel but-
tons provide the user with a rich feature set. The 8800 Series Trace Moisture Analyzer's advanced
design allows it to be housed in a variety of enclosures.
The 8800 Series has an impressive set of dewpoint measurement capabilities in terms of accuracy,
stability, response time etc. The specifications of the sensor are discussed in section 2.2, while the
specifications of the instrument are discussed in section 3.2. A summary of the standard and
optional features & capabilities of the 8800 Series are listed below as an overview aid to the user.
Standard Features/Capabilities
Locking of instrument, preventing unintentional changes
SpanCheck™: - automatic recalibration using room air, all instruments pre-calibrated at fac-
tory.
Capability to enter up to 15 NIST/NPL traceable calibration points at factory, depending on
order.
Manual Self Calibration: a single point calibration using a known standard gas. High acccu-
racy can be maintained even in the most harsh applications.
Pressure correction: built-in software calculation of dewpoint at a pressure different than the
measurement.
Cable length compensation: automatic self measuring software.
Universal autoranging AC supply 100-250VAC
Instrument and Sensor: UL & cUL listed/recognized; CE electromagnetic compatibility certi-
fied
Optionally ordered Features/Capabilities
Password protected Locking of instrument, preventing unauthorized changes
Alarm Relays - programmable set points, error handling, and hysteresis. Up to 3 alarm relays
may be ordered. On-display legends indicate relay states.
Audio Visual Alarm- complies with recommendations of NFPA 99, 1996 edition, audio reset
and test buttons.
Analog Output- factory set voltage or user selectable 4/20 - 0/24 mA. Includes installation
testing features. User selectable range (low & high point) of analog output.
RS-232C interface - allowing the user digital data access to the instrument.
Interval-timer-scheduled Self Calibration, a single point calibration using a known standard
gas, can operate an electrically actuated switchover valve for unattended calibrations
Large variety of threads for sensor mounting into sample
Low voltage DC power operation 15-30VDC
Intrinsically Safe Approved NEC and CENELEC standards (UL & DEMKO) configurations
1