Part I: Control Unit
Aside from normal cleaning and checking for leaks at the gas connec-
tions, routine maintenance is limited to replacing Micro-Fuel cells and fuses,
and recalibration.
Checking for leaks, replacing Micro-Fuel cells, and replacing fuses in
the Analysis Unit are covered in Part II, Chapter 5. For recalibration, see
Part I, section 4.4 Calibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
5.1
The 3010TB requires two 5 x 20 mm, 1.0 A, T type (Slow Blow)
fuses.
The fuses are located inside the main housing on the Electrical
Connector Panel, as shown in Figure 5-3. To replace a fuse:
Maintenance
MANUAL.
Fuse Replacement
1. Disconnect the Unit from its power source.
2. Place a small screwdriver in the notch in the fuse
holder cap, push in, and rotate 1/4 turn. The cap will
pop out a few millimeters. Pull out the fuse cap and
fuse, as shown in Figure 5-1
Teledyne Analytical Instruments
Maintenance 5
Part I: 5-1