Figure 7-10. Ground Bond Resistance Calibration on Fluke 6500 Using Single Test Leads
Figure 7-11. Ground Bond Resistance Calibration on Bench Ground Bond Tester

Calibrating Line Impedance Testers

The Calibrator's Line Impedance calibration function calibrates the line impedance
function of loop testers and multifunction installation testers with loop testing capability.
Different models of installation testers use different test current levels to avoid tripping
protective circuits. The Calibrator limits the amount of test current that can be used
during a line impedance calibration.
To avoid electric shock, do not touch the L, PE, or N terminals
on the Calibrator or UUT while performing a Line Impedance
calibration. Line voltage is present on these leads during this
calibration.
To perform a Line Impedance calibration:
7-12
WX Warning
Fluke 5320A
5320A MULTIFUNCTION ELECTRICAL TESTER CALIBR
OUTPUT
V
, HI
, mA
LO
LO
- SENSE
HI
HI
50V PK
1500V PK
MAX
MAX
LO
LO
20V PK
20V PK
Z
, Z
GND ,
RCD
L
L
PE
N
L1
H
L2
E
L3
S
280V
280V
RMS
RMS
MAX
MAX
20V PK
METER
INPUT
METER
A
COM
V
CAT I
30A
1000V
RMS
CAT II
20V PK
600V
MAX
RMS MAX
ehq035.eps
Fluke 5320A
5320A MULTIFUNCTION ELECTRICAL TESTER CALIBR
OUTPUT
V
, HI
, mA
LO
LO
- SENSE
HI
HI
50V PK
1500V PK
MAX
MAX
LO
LO
20V PK
20V PK
Z
, Z
GND ,
RCD
L
L
PE
N
L1
H
L2
E
L3
S
280V
280V
RMS
RMS
MAX
MAX
20V PK
METER
INPUT
METER
A
COM
V
CA T I
30A
1000V
RMS
CAT II
20V PK
MAX
600V
RMS MAX
ehq055.eps

Advertisement

Table of Contents
loading

Table of Contents