EFI DocuColor 5252 Supplementary Manual page 83

Color printing
Table of Contents

Advertisement

C
M
OLOR
EASUREMENT
9 If you have not already done so, remove the shipping spacer from the measurement
page slot.
10 Insert the end of the Color Reflection Reference strip with the arrow into the calibration strip
entrance on the DTP41, centering it below the alignment mark (see the diagram on
Insert the strip about 3 inches, past the front idler rollers, until it rests against the rear
drive rollers.
11 Click Read Reference.
The DTP41 pulls the calibration strip through automatically.
A message appears in the Status area when calibration is finished.
12 Click Done to close the Instrument Calibration dialog box.
13 Close the Utilities dialog box.
This completes DTP41 calibration.
Your computer prompts you when the DTP41 requires calibration. When the prompt
message appears, click Calibrate Now and follow the instructions, starting with
page
82.
After calibrating the DTP41, calibrate the Fiery EXP6000/EXP5000 (see
I
NSTRUMENTS
83
page
79).
step 8
on
page
47).