Non-Latching Relay (Auto); Non-Latching Relay (A-Man); Latching Relay (Latch) - Omega DP6000 User Manual

Dual-line process meter
Table of Contents

Advertisement

DP6000 Process Meter

Non-Latching Relay (Auto)

Automatic reset only
Condition
Normal
Alarm
Ack (No effect)
Normal
In this application, the meter is set up for automatic reset (non-latching
relay). Acknowledging the alarm while it is still present has no effect on
either the LED or the relay. When the alarm finally goes away, the relay
automatically resets and the LED also goes off.

Non-Latching Relay (A-man)

Automatic + manual reset at any time
Condition
Normal
Alarm
Normal
Next Alarm
Ack
Normal
In this application, the meter is set up for automatic and manual reset at
any time (non-latching relay). The LED and the relay automatically reset
when the meter returns to the normal condition.
The next time an alarm occurs, the operator acknowledges the alarm
manually while the alarm condition still exists. This causes the relay to
reset, but the LED stays on until the meter returns to the normal
condition.

Latching Relay (LatcH)

Manual reset any time
Condition
Normal
Alarm
Ack
In this application, the meter is set up for manual reset at any time.
Acknowledging the alarm even if the alarm condition is still present
resets the relay and turns off the LED.
LED
Relay
Off
Off
On
On
On
On
Off
Off
LED
Relay
Off
Off
On
On
Off
Off
On
On
On
Off
Off
Off
LED
Relay
Off
Off
On
On
Off
Off
60
Instruction Manual

Advertisement

Table of Contents
loading

Table of Contents