Commissioning
7.4 Description of OSD menu
7.4.3

Lock/unlock OSD menu

CAUTION
Locking/unlocking of the OSD is only permissible for authorized servicing personnel. The
OSD must be locked if a faulty operation on the part of the user could have a detrimental
effect on the approved application of the monitor.
Lock
You can lock the call from the OSD if the OSD is not active.
To lock, enter the following key combination without interruption:
● Press button "4" once and then button "2" three times in quick succession.
The OSD menu is locked.
Cancel locking
● Press button "4" once and then button "2" three times (if the OSD is not active).
Locking of the OSD menu has been canceled.
Delivery state
The OSD is unlocked.
7.4.4

Picture layout (PaP/PiP/PoP)

PaP: Picture and picture
Picture and picture" layout; two images from main and secondary image sources are
displayed next to each other.
PiP: Picture in picture
"Picture in picture" layout; the image content of a secondary window source is displayed over
the main picture (≙ main window source).
PoP: Picture on picture
"Picture on picture" layout; up to five images from different image sources are displayed next
to each other.
● The image from the main image source is displayed larger on the bottom side of the
window.
● The other images are displayed in a smaller format on the top side.