Maintenance - Eizo RadiForce LX470W Instruction Manual

2 mp 47" color flat panel monitor
Table of Contents

Advertisement

10.2

Maintenance

Checking the monitor settings at regular intervals
The monitor is low-maintenance. The illumination properties of the monitor may change due
to aging of the LCD unit and the backlight.
● Check the monitor settings against the country-specific guidelines at regular intervals.
Measuring instruments
The following devices can be used as measuring instruments:
● Serial Spot Meter
● Universal Serial Luminance Meter
● Advanced Serial Luminance Meter
These photometers can be connected directly to the monitor.
Carrying out quality tests automatically
All quality tests can be performed automatically by the EIZO QA software.
Verifying and calibrating device properties
You can verify the device properties and calibrate them if necessary using the photometers.
Confirming the image quality visually after calibrating the monitor
After calibration, the monitor must be visually inspected to verify successful and correct
completion of the calibration procedure. This inspection is performed by reviewing the
SMPTE image.
● The gray levels must be displayed correctly and visibly at both 5% and 95%.
● Alternatively, the VerLum image can serve as a useful test pattern.
Successful calibration can be verified if the small squares are displayed correctly in all
gray shades.
See also
Check for pixel defects (Page 50)
Program for changing the display settings (Page 51)
RadiForce® LX470W