5.3.2. Laser sensor module
When
Estimated time to complete [min.]: 30
Required special tools
Required part(s)
1. Prerequisites
•
Contact your local service engineer.
•
Order a NEW laser sensor unit.
NOTE: It is not allowed to open, trouble shoot or
repair parts of the Laser sensor unit in or outside the
AX-201. All warranty will expire.
2. Remove the Sensor Unit
•
Switch off and power down the AX-201
•
Lock the main switch with a pad-lock
•
Disconnect the Sensor connection cable
and the safety extension cable.
•
Loosen the two bolds.
•
Remove the unit from the base
3. Place the new sensor unit
•
Place the new unit completely over the
dowel pins
•
Fasten the two bolds.
•
Check the serial number in the
'settings.xml'. See Chapter 3.3.2.2
•
Connect the sensor cable to the sensor.
•
Connect the safety cable to the safety
extension cable.
•
Power up the AX-201.
Follow the teach X, Y coordinates
procedure (See chapter 3.3.2.5)
If the laser sensor is not working
N.A.
New Laser sensor unit (9466-916-00051)
39