5. Maintenance and Service
5.1. Maintenance
Note: No maintenance other than mentioned in this manual may be performed on the Coplanarity laser
sensor. If done otherwise; all warranty will be expired.
5.1.1. Glass plate cleaning
When
Estimated time to complete [min.]: 10
Required special tools
1. Prerequisites
•
Stop production.
•
Power down the AX-201
•
Close the shutter on the sensor
2. Clean the sensor window
•
Clean the sensor window with a small
amount of isopropanol and a soft cloth
Remark:
Make sure no residue is left on the glass plate.
Residue can affect measurements.
3. Prepare for production
•
Open the shutter
•
Power up the AX-201.
36
Once every 6 month
Soft cloth, isopropanol, lens tissues