Sequence of Operations
Refer to the table of UUT calibration points in the UUT Manufacturers Calibration Guide.
Follow the correct sequence of calibration stages as directed by the guide, and carry out
the following operations (1) to (5) at each stage.
1. 9100
a. Use the front panel controls to set the 9100 Output to the UUT
cal point frequency, voltage and current, entering High Voltage
State if the cal point has been assigned to that State. The default
High Voltage State boundaries are shown in Fig. 4.20.1.
b. Select the required waveforms and phase-shift.
2. UUT
Select the correct range for the cal point.
3. 9100
a. Set Output ON.
b. Slew the AC Power Output reading until the UUT reading is
equal to the calibration point value.
4.
UUT
Record the 9100 screen output value as detailed in the UUT
Manufacturer's Calibration Guide.
5.
9100
Set Output OFF.
Section 4: Using the Model 9100: AC Power Function - Operation