3. On the Honeywell Versatilis Configurator app, go to the Loop Test method..
4. Select the desired fixed output level: 0%, 100%, or Other (any between 0% - 100%) and proceed.
5. A warning will display to confirm placing the transmitter in fixed output mode.
With the transmitter in Analog mode, you can observe the output on an externally-
connected meter or on a local meter. In DE mode, you can observe the output on the local
meter or on the Configurator display.
6. Observe the output current at the percentage you selected in Step 5.
7. Exit Loop Test and view the monitored PV.
8. This concludes the Startup procedure.