86
Warning: this DD is untested with the
current Hart App ver <manufacturer
name, device name, dev. rev. #, and
DD rev. #>...you may experience
unexpected or inappropriate
results...See Programming Interface
for details on DD updates...Do you
want to accept these risks and
proceed?
DD Error..."tested-to sequence
mismatch"...DD version
<manufacturer name, device name,
dev. rev. #, and
DD rev. #>; HART application
<version x.y>; HART app version
x.y...See Programming Utility for
solution.
DD Error..."developed-to
mismatch"...DD version
<manufacturer name, device name,
dev. rev. #, and
DD rev. #>; HART application
<version x.y>; HART app version
x.y...See Programming Utility for
details
Device Description Error...'Device
Description header file
corrupted'...To repair, reinstall the
Device Descriptions for %s %s dev
rev %d DD rev %d...see
Programming Utility for
details...Press FWDCOM to attempt
to load an earlier Device
Description."
Communication Error...See User's
Manual for solution.
Troubleshooting
The device description was not
tested and verified with the
installed version of the HART
Application. Obtain a newer
version of the device description
using the Field Communicator
Easy Upgrade Utility.
The DD was tested with a previous
version, but not with this version of
the HART Application. Obtain a
newer version of the device
description using the Field
Communicator Easy Upgrade
Utility.
The DD was developed using a
later HART Application revision.
Upgrade the system software and
device descriptions using the Field
Communicator Easy Upgrade
Utility.
There is a checksum error in the
device description header file.
Obtain a newer version of the
device description using the Field
Communicator Easy Upgrade
Utility.
Check the wiring to the device.