54
To change the polling option:
1.From the HART Application main menu, double-tap Utility.
2.Double-tap Configure HART Application.
3.Double-tap Polling Options. Select one of the following polling
options:
• Poll By Address - Lets you poll for devices at the specified
addresses. Tap Enter and then double-tap Polling Addresses
from the Configure HART Application window to select addresses
to poll. (This option appears only when Poll By Address is
selected.) Select one of the following options from the Polling
Addresses menu:
Table 3-2. Options in the Polling Addresses menu
Polling
Recommended Use
Addresses
0
Poll non multi-dropped loops.
0-15
Poll HART Universal Revision 5 or earlier devices
on multi-dropped loops.
0-63
Poll HART Universal Revision 6 or newer devices
on multi-dropped loops.
63
Poll the default address for WirelessHART
adapters.
Custom Range
Poll a unique or non-consecutive group of
addresses (example: "0, 5-7, 12"). Use a hyphen
to specify consecutive addresses and/or a comma
to separate numbers or groups of numbers.
• Poll Using Tag - Lets you enter and poll by the tag of the device.
You are prompted to enter a tag when the HART Application
starts.
• Poll Using Long Tag - Lets you enter and poll by the long tag of
the device. (Only supported in HART Universal Revision 6 or
higher devices.) You are prompted to enter a tag when the HART
Application starts.
• Poll Using Unique ID - Lets you poll using the device's unique ID.
For HART Universal Revision 7 devices, the unique ID is the
expanded device type, a hyphen, and the device ID. For HART
Universal Revision 6 and earlier devices, the unique ID is the
manufacturer ID, the device type, a hyphen, and the device ID.
You are prompted to enter a unique ID when the HART Application
starts.
4.Tap Enter.