Appendix A
Appendix A

General Safety Information

This is a Safety Class I instrument (provided with terminal for protective
earthing) and has been manufactured and tested according to international
safety standards.
This instrument is designed as a laboratory equipment. Use it in an analytical
WA R N I N G
environment only.
Use this instrument in a manner described in this manual.

Operation

Before applying power, comply with the installation section. Additionally the
following must be observed.
Do not remove instrument covers when operating. Before the instrument is
switched on, all protective earth terminals, extension cords,
auto-transformers, and devices connected to it must be connected to a
protective earth via a ground socket. Any interruption of the protective earth
grounding will cause a potential shock hazard that could result in serious
personal injury. Whenever it is likely that the protection has been impaired,
the instrument must be made inoperative and be secured against any intended
operation.
Make sure that only fuses with the required rated current and of the specified
type (normal blow, time delay, and so on) are used for replacement. The use of
repaired fuses and the short-circuiting of fuseholders must be avoided.
Any adjustment, maintenance, and repair of the opened instrument under voltage is
WA R N I N G
forbidden.
Disconnect the instrument from the line and unplug the power cord before
WA R N I N G
maintenance.