Arm the Timing Analyzer with the State Analyzer

Arming control sets up the order of triggering for complicated measurements
that involve more than one mode, such as state and timing. Although state
and timing modes involve a single analyzer, they are often referred to as the
timing analyzer and the state analyzer. Arming the timing analyzer with the
state analyzer allows the two modes to run simultaneously so that the data
captured by both analyzers can be time-correlated.
Highlight the
1
If you are using an HP 1670 series logic analyzer, the Arming Control pop-up
menu will look different because these machines do not have an oscilloscope.
Highlight
2
field. Press the
Highlight the field next to
3
Highlight
4
key. Press the
Press the Done key to return to the Trigger Menu.
5
Machine 1 is the
state analyzer.
Machine 2 is the timing
analyzer.
Arming the Timing Analyzer with the State Analyzer
Result
When the state analyzer (Machine 1) arms the timing analyzer (Machine 2),
the timing analyzer begins to look for its own trigger condition.
Mixed Mode - Correlation of State and Timing Data
Arm the Timing Analyzer with the State Analyzer
field, and press the
Arming Control
MACHINE 2
near the center of the screen, next to the
key.
Select
, and press the
Run from
in the pop-up menu, and press the
MACHINE 1
Done
key to close the Machine 2 pop-up menu.
Get other manuals https://www.bkmanuals.com
key.
Select
Run
key.
Select
Select
The timing analyzer is
armed by the state
analyzer.
5-9