After Use; Disconnecting The Co2 Gas Cylinder - Olympus UCR Instructions Manual

Endoscopic co2 regulation unit
Table of Contents

Advertisement

5.5 After use

5.5
After use
Disconnecting the CO
Ch.5
1
Close the gas cylinder valve by turning it
clockwise.
2
Disconnect the water container-side connector of the gas tube from the gas tube
connector on the water container (MAJ-902).
3
Disconnect the Gas tube (MAJ-1741), Low flow gas tube (MAJ-1742), or Extra low
flow gas tube (MAJ-1816) from the CO
unit.
4
Press the start/stop switch of this instrument to start the gas insufflation for removing
CO
gas remaining inside the endoscopic CO
2
5
After the endoscopic CO
6
Disconnect the power cord plug from the hospital grade AC outlet (wall mains outlet).
50
WARNING
When disconnecting the high-pressure hose from the gas supply port on the rear
panel and the CO
cylinder, be sure to close the CO
2
CO
remaining in the instrument. Otherwise, a large amount of gas will jet out of
2
the instrument.
When disconnecting the endoscopic CO
CO
gas inlet and CO
2
closed, then release any residual CO
regulation unit. The gushed high-concentration CO
difficulty in breathing if inhaled.
2
regulation unit has entered the stop mode, turn it OFF.
2
UCR INSTRUCTION MANUAL
regulation unit's cylinder hose from the
2
gas cylinder, ensure that the CO
2
gas remaining inside the endoscopic CO
2
gas cylinder
CO
cylinder
Figure 5.4
gas outlet of the endoscopic CO
2
regulation unit.
2
cylinder valve and exhaust
2
gas cylinder' valve is
2
is hazardous and may cause
2
Gas cylinder
valve
gas
2
regulation
2
2