Chapter 5 Operation
5.9

Procedure after use

56
Always discard used disposable patient plates. Never
attempt to reprocess or reuse them. It may cause patient
burns.
Do not use the P cord as a tool to remove the patient plate
from the patient. If utilized in this fashion, skin injuries can
occur.
When using the disposable patient plate
When disconnecting the plug on the unit side of the P cord
(MAJ-814) or patient plate, always hold the plug on the unit
side or patient plate clamp. Pulling the cable may result in
damaging of the wires.
1.
Turn the electrosurgical unit OFF.
2.
Remove the patient plate from the patient and disconnect the patient plate
plug from the patient connector on the front panel.
3.
Disconnect the A-cord's plug on the accessory side from the accessory for
the endoscope. Disconnect the A-cord's plug on the unit side from the active
connector on the front panel.
4.
Following endoscopic treatment, disconnect the S-cord's plug on the
endoscope side from the endoscope's S-cord connector mount. Pull the ring
on the S-cord's plug outward and disconnect the S-cord from the S-cord
connector on the front panel (for an endoscope equipped with the S-cord
connector mount).
5.
When the foot switch is no longer needed, rotate the fastener ring on the
foot switch plug counterclockwise and disconnect the plug from the foot
switch connector on the rear panel.
6.
If the electrosurgical unit is not to be used for a long period, disconnect the
power cord plug from the wall mains outlet.
ELECTROSURGICAL UNIT PSD-30