TIMebase Subsystem

RANGe

RANGe
Command
:TIMebase:RANGe <range>
The RANGe command sets the full-scale horizontal time in seconds. The
RANGE value is ten times the value in the s/Div field.
<range>
time in seconds
Example
OUTPUT XXX;":TIMEBASE:RANGE 2US"
Query
:TIMebase:RANGe?
The RANGe query returns the current setting.
Returned Format
[:TIMebase:RANGe] <range><NL>
Example
OUTPUT XXX;":TIMEBASE:RANGE?"
8–6