RANGe

Command
:CHANnel<N>:RANGe <range>
The RANGe command defines the full-scale (4 times the Volts/Div) vertical
axis of the selected channel. The values for the RANGe command are
dependent on the current probe attenuation factor for the selected channel.
The allowable range for a probe attenuation factor of 1:1 is 16 mV to 40 V.
For a larger probe attenuation factor, multiply the range limit by the probe
attenuation factor.
An integer from 1 through the number of channels in the oscilloscope
<N>
connected as one unit (maximum of 8 channels), otherwise the maximum
number of channels is 2.
16 mV to 40 V for a probe attenuation factor of 1:1
<range>
Example
OUTPUT XXX;":CHANNEL1:RANGE 4.8"
Query
:CHANnel<N>:RANGe?
The RANGe query returns the current range setting.
Returned Format
[:CHANnel<N>:RANGe] <range><NL>
Example
OUTPUT XXX;":CHANNEL1:RANGE?"
CHANnel Subsystem
RANGe
4–9