Ai Pause Trigger Signal - National Instruments X Series User Manual

Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Chapter 4
Analog Input

AI Pause Trigger Signal

Use the AI Pause Trigger (ai/PauseTrigger) signal to pause and resume a measurement
acquisition. The internal sample clock pauses while the external trigger signal is active and
resumes when the signal is inactive. You can program the active level of the pause trigger to be
high or low, as shown in Figure 4-24. In the figure, T represents the period, and A represents the
unknown time between the clock pulse and the posttrigger.
Figure 4-24. Halt (Internal Clock) and Free Running (External Clock)
AI Sample Clock
AI Convert Clock
AI Pause Trigger
AI External Sample Clock
AI Sample Clock
AI Convert Clock
AI Pause Trigger
Using a Digital Source
To use AI Pause Trigger, specify a source and a polarity. The source can be any of the following
signals:
PFI <0..15>
RTSI <0..7>
PXI_STAR
PXIe_DSTAR<A,B>
Counter n Internal Output
Counter n Gate
AO Pause Trigger (ao/PauseTrigger)
DO Pause Trigger (do/PauseTrigger)
DI Pause Trigger (di/PauseTrigger)
The source can also be one of several other internal signals on your DAQ device. Refer to Device
Routing in MAX in the NI-DAQmx Help or the LabVIEW Help for more information.
4-34 | ni.com
A
T
Halt. Used on Internal Clock
Free Running. Used on External Clock
T – A

Advertisement

Table of Contents
loading

Table of Contents