The timing engine ignores the AI Start Trigger signal while the clock generation is in progress.
After the clock generation is finished, the counter waits for another Start Trigger to begin another
clock generation. Figure 4-22 shows a retriggerable analog input with three AI channels and four
samples per trigger.
AI Start Trigger
AI Sample Clock
AI Convert
Note
Waveform information from LabVIEW does not reflect the delay between
triggers. They are treated as a continuous acquisition with constant t0 and dt
information.
Reference triggers are not retriggerable.
Using a Digital Source
To use AI Start Trigger with a digital source, specify a source and an edge. The source can be
any of the following signals:
PFI <0..15>
RTSI <0..7>
Counter n Internal Output
PXI_STAR
PXIe_DSTAR<A,B>
Change Detection Event
AO Start Trigger (ao/StartTrigger)
DI Start Trigger (di/StartTrigger)
DO Start Trigger (do/StartTrigger)
The source can also be one of several other internal signals on your DAQ device. Refer to Device
Routing in MAX in the NI-DAQmx Help or the LabVIEW Help for more information.
You can also specify whether the measurement acquisition begins on the rising edge or falling
edge of AI Start Trigger.
Using an Analog Source
When you use an analog trigger source, the acquisition begins on the first rising edge of the
Analog Comparison Event signal.
Figure 4-22. Retriggerable Analog Input
X Series User Manual