EAN/UPC Calibration Cards:
1. Choose the appropriate "Master Grade" barcode. The "EAN-13 Master Grade"
code is used to calibrate systems in Europe. The "UPC-A Master Grade" code
is used to calibrate systems within the United States.
2. Place the calibration card in the system's field of view.
3. Choose one of the "Master Grade" barcodes and place it so the blue line goes
through the middle of the "PASS" portion of the barcode.
4. Confirm that the "Goal" values in the software (on the "Calibration" tab) match
the values on the calibration card. If the values do not match, confirm that you
have the correct calibration card for the LVS-95XX system serial number
(located on system) that you are calibrating.
GS1-128 Calibration Cards:
Larger fields of view require a "GS1-128" calibration card. Use the "PASS" portion
of the GS1-128 barcode to calibrate your system.
Data Matrix Calibration Card for ISO/IEC Data Matrix:
This Calibration Test Card is used only for calibrating the LVS-9585-DPM-HD and
the LVS-9580-DPM-HD systems.
1. Use Application Standard ISO/IEC 15415/15416 when calibrating with this
card.
2. Place Symbol 1 on the Calibration Card in the blue box located on the
Calibration Screen.
LVS-95XX Frequently Asked Questions
DATA MATRIX Conformance
Calibration Standard Test Card
Page 14 of 17