Inspection and maintenance
PMF
kV and tube current (IR)
Remove the 4 Allen screws, and then remove the upper control box cover.
Connect the oscilloscope:
CH1: D915.IR
CH2: D915.kV (1 V 20 kV actual value)
and D915.GND
Connect the power plug, POLYMOBIL ON.
Trigger exposure with 90 kV, 10 mAs.
If the oscillogram ist not produced, perform an adjustment according to the service
instructions.
POLYMOBIL OFF and disconnect the power plug.
Close the control console.