Inspection and maintenance
PMF
kV and tube current (IR)
•
Remove the 4 Allen screws, and then remove the upper control box cover.
•
Connect the oscilloscope:
CH1: D915.IR
CH2: D915.kV (1 V 20 kV actual value)
and D915.GND
•
Connect the power plug, POLYMOBIL ON.
•
Trigger exposure with 90 kV, 10 mAs.
•
If the oscillogram ist not produced, perform an adjustment according to the service
instructions.
•
POLYMOBIL OFF and disconnect the power plug.
•
Close the control console.