Fig. 2 Measuring circuit for measuring the protective conductor resistance in systems that are disconnected from
the power supply, according to DIN VDE 0751-1/2001-10, Fig. C2.
1 = System
2 = Applied part, type B (if present)
3 = Measuring arrangement (integrated in the measuring instrument)
The values determined must be recorded and assessed in the protective conductor resis-
tance report and include the measured points.
Document the measuring procedure and the measuring instrument used (designation
and serial number).
NOTE
NOTE
Siemens AG
Medical Solutions
The protective conductor resistance report is filed in the
"Certificates" register in the system binder.
A new report must be created if the protective conductor resis-
tance measurements are not documented.
These instructions contain a new report called 'Protective con-
ductor resistance/report'. Separate this report, fill it out, and file
it in the system manual under "Certificates". If values are newly
determined, they are to be recorded as first measured values.
The evaluation is omitted in this case.