Fig. 3 Measuring circuit for measuring the system leakage current using the differential current method
according to DIN VDE 0751-1/2001-10, Fig. C6 for protection class
1 = System
2 = Applied part, type B (if present)
3 = Measuring arrangement (integrated in the measuring instrument)
If the direct measurement of the system leakage current is used (measuring arrangement
according to Fig. 4), then the system must be set up insulated during the measurement
and must not be touched during the measurement.
Fig. 4 Measuring circuit for the direct measurement of the system leakage current
according to DIN VDE 0751-1/2001-10, Fig. C5 for protection class
1 = System
2 = Applied part, type B (if present)
3 = Measuring arrangement (integrated in measuring instrument)