Display Menu; Display Setting; Figure 13 Display Menu - Fluke 1620A-H User Manual

Thermo-hygrometer
Hide thumbs Also See for 1620A-H:
Table of Contents

Advertisement

7 Menu Functions

Display Menu

rameters have been edited. The message 'STORE VALUES' is displayed. Press
the Enter button to save the new settings. The values are stored in the sensor.
Press the Exit button to cancel any changes and exit to the menu.
7.2
Display Menu
The DISPLAY menu provides functions for configuring the display. The func-
tions that appear in this menu are DISPLAY SETTING, DISPLAY LAYOUT,
FIELD DATA, GRAPH SCALE, and DISPLAY RESET.

Figure 13 Display Menu

7.2.1

Display Setting

The DISPLAY SETTING function allows you to select the temperature display
resolution (T RES), humidity display resolution (H RES), decimal preference,
unit of temperature and display contrast. These settings may be password-pro-
tected (see Section 7.5.4, Password).
The temperature and humidity display resolutions are selectable between 0, 0.1,
0.01, and 0.001. These settings do not affect the resolution of recorded or
printed data. Recorded and printed data are fixed at 0.01°C or °F and 0.1% RH.
The decimal preference is selectable between a period (.) and a comma (,).
The available units of measure are degrees C and degrees F.
The contrast can be adjusted here using the lr buttons. Setting the adjust-
ment number more negative lightens the screen and setting it more positive
31

Advertisement

Table of Contents
loading

This manual is also suitable for:

1620aDewk1620a-s

Table of Contents