The <bool> parameter turns the measurement averaging on (1 or ON) or off (0
or OFF). This command may be password protected (see Section 8.2.10, Pass-
word Commands).
8.2.4.5
SENSor<chn>:LOCK?
Returns the sensor lock state for the given channel number.
Example: SENS1:LOCK?
Response: 1
A value of 1 is returned if the sensor lock is enabled. A value of 0 is returned if
the sensor lock is disabled.
8.2.4.6
SENSor<chn>:LOCK <bool>
Sets the sensor lock for the specified channel.
Example: SENS1:LOCK 1
The <bool> parameter turns the sensor lock on (1 or ON) or off (0 or OFF).
This command is password protected. To access this command when the
SYST:PASS:SENS is enabled, the password must be successfully entered using
the SYST:PASS:CEN <pass> command (see Section 8.2.10, Password
Commands).
8.2.4.7
SENSor<chn>:IDENtification?
Returns the sensor identification for the specified channel.
Example: SENS1:IDEN?
Response: "LABORATORY"
The SENSor suffix, <chn>, specifies the channel (1 or 2). The response to this
command is enclosed in double quotes and indicates the current ID string for
the sensor on the specified channel.
8.2.4.8
SENSor<chn>:IDENtification <str>
Sets the sensor identification for the specified channel.
Example: SENS1:IDEN CHAMBER3
The SENSor suffix, <chn>, specifies the channel (1 or 2). The sensor identifi-
cation field is set to the value indicated by <str>. The sensor identification can
be up to 16 characters in length and consist of any combination of numbers, let-
ters, underscores, and spaces.
This command may be password protected (see Section 8.2.10, Password
Commands)
8 Digital Communications Interface
Commands
103