The <bool> parameter turns the alarm beep enable on (1 or ON) or off (0 or
OFF). This command may be password protected (see Section 8.2.10, Pass-
word Commands).
8.2.1.6

ALARm:CLEar

Clears the alarm events.
Example: ALAR:CLE
The alarm clear command clears the alarm events.
8.2.1.7

ALARm:DATE:FIRSt?

Returns the date of the first alarm event.
Example: ALAR:DATE:FIRS?
Response: 2003,8,10
The response is returned in the format, <year>,<month>,<day>.
If no alarm has been triggered, the response is: 2000,0,0.
8.2.1.8

ALARm:DATE:LAST?

Returns the date of the last alarm event.
Example: ALAR:DATE:LAST?
Response: 2003,8,10
The response is returned in the format, <year>,<month>,<day>.
If no alarm has been triggered, the response is: 2000,0,0.
8.2.1.9

ALARm:DISPlay:ENABle?

Returns the state of the visible alarm enable.
Example: ALAR:DISP:ENAB?
Response: 0
A value of 1 is returned if the alarm display is enabled. A value of 0 is returned
if the alarm display is disabled.
8.2.1.10
ALARm:DISPlay:ENABle <bool>
Sets the state of the visible alarm enable.
Example: ALAR:DISP:ENAB 1
The <bool> parameter turns the alarm display enable on (1 or ON) or off (0 or
OFF). This command may be password protected (see Section 8.2.10, Pass-
word Commands).