Controlling The System With E8363/4C And N5230C; Pna Multiport Mode With Option 551 - Agilent Technologies U3025AE10 User Manual

Hide thumbs Also See for U3025AE10:
Table of Contents

Advertisement

U3025AE10

Controlling the System with E8363/4C and N5230C

This section will describe how to setup and operate the U3025AE10 Multiport Test Set
with the E8363/4C and N5230C 2-Port PNA Network Analyzer.
The U3025AE10 Multiport Test Set is considered a "slave" instrument. A E8363/4C or
N5230C must be used to control the Test Set.
• PNA Multiport Mode Option 551 requires firmware revision ≥ A.08.04.10. From the
PNA drop-down menu select Help > About Network Analyzer to see the options
installed and the installed firmware revision. See
The following test set file must be installed into network analyzer file directory:
c:\Program Files\Agilent\Network Analyzer\testsets
2-Port PNA-X requires test set: u302xae10_pnax_p2.tsx
2-Port PNA or PNA-L requires test set: u302xae10_p2.tsx
Visit our website for firmware revision and
downloads.http://na.tm.agilent.com/pna/firmware/firmware.htm
You may also refer to

PNA Multiport Mode with Option 551

The PNA Multiport Mode selects the test set file that will enable the PNA to control the
Test Set. The PNA Multiport Mode allows you to complete a N-Port calibration using the
Cal Wizard application in the PNA. Refer to the PNA Help system for more information.
Figure 24
N5230C and E8363/4C Options
The internal firmware of the Agilent E8363/4C and N5230C has not been
NOTE
modified for this test set option. Power levels may differ from those indicated
on the PNA when the Test Set is connected.
User's Guide
"Installing New Firmware" on page 20
Controlling the System with E8363/4C and N5230C
Figure
24.
in this manual.
29