Adjustments

Image Rotation

Fig. 2:
Fig. 3:
Image_rotation
Adjustment
1. Register an emergency patient to allow radiation release.
2. Manually set the image rotation to the 0 degree position at the control console.
3. Attach an Allen key and a piece of solder wire to the image intensifier as shown in
(Fig. 3 / p.
4. Click on <Go>.
X
5. Briefly release fluoro and then use the PC mouse to rotate the LIH image on the moni-
tor screen into the 0 degree position.
6. Click on <Go>.
Siemens AG
Medical Solutions
Image rotation
See display for image rotation at the control console.
11).
Status bar: Init (Disable Limits) --> running --> success
The image (left/right / up/down) corresponds to
Status bar: 0 Degree Position --> running --> success
SPR2-310.842.01.07.02
12.06
Fig. 4:
Image position
(Fig. 4 / p.
Page 11 of 34
CS SD 24
11
11).