12. Briefly release fluoro for the LIH image.
13. Adjust the display lines on the monitor until the lines are aligned with the front edges of
the slot diaphragm blades.
14. Click on <Go>.
Init Off & Store Position
ARCADIS Varic / Orbic
A window appears: "Calibration is finished..." --> confirm with <OK>.
The main system restarts.