Image intensifier tube

Fig. 7
Detach the camera attachment (5/Fig. 4); lead.
Remove the I.I. tube from the I.I. housing.
CAUTION
Completely dispose of the lead clad housing of the image intensifier.
Vent the I.I. tube by pinching it open at the pump port (1/Fig. 7).
Substances at the input screen (2/Fig. 7)
Substances at the output screen (3/Fig. 7)
Substances at the ceramics part (4/Fig. 7)
UROSKOP ACCESS
System
Danger of implosion!
Avoid mechanical damage to the I.I. tube.
SPL5-330.861.01
Rev. 02
04.05
Disassembly and Disposal
CS/NF/93/290
CSJ/NaJ
Sb
Cr
O
2
3
phosphorus NP 1043
lead-containing solders
silicone sealing compound
Cr
O
2
3
Page 8 of 8