Adjustment system
Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

12
7. Click on <Go>.
8. Manually set the image rotation to the 0 degree position at the control console.
X
9. Briefly release fluoro.
10. Click on ">" for the next page and click on <Finish>.
ARCADIS Varic / Orbic
Status bar: Init Off & Store Positions --> running --> success A window appears:
Calibration is finished --> confirm with <OK>.
The main system restarts.
See display for image rotation at the control console.
The image should now display in the correct 0 degree position
not, repeat adjustment.
SPR2-310.842.01.07.02
12.06
Page 12 of 34
CS SD 24
Adjustments
(Fig. 4 / p.
11); if
Siemens AG
Medical Solutions