5.2 Various problems and countermeasures

5 .2 .5 Detector Interface Unit 2

Status
LED(green/blue) does not light.
LED (blue) does not light.
The DR Detector is being used
with wired connection, but it is
not recognized on the image
processing controller.
Serial exposure does not start.

5 .2 .6 Power Supply Unit

Status
LED does not light.
The power switch cannot be
turned on.
The DR Detector is being used
with wired connection, but it is
not recognized on the image pro-
cessing controller.

5 .2 .7 AeroDR Generator Interface Unit

Status
LED (green) does not light.
LED (green) does not change
from flashing to a solid light.
Error description
AC power not supplied.
The I/F cable is not connected to the DR
Detector.
The I/F cable is not connected to the DR
Detector. Or, the Ethernet cable is not
connected to the Power Supply Unit.
The Ethernet cable for relaying serial exposure
signals is not connected to the Sync port.
Error description
AC power not supplied.
There is an error in the power system.
The Ethernet cable is disconnected.
Error description
AC power not supplied.
There is a communications error.
Make sure that the Ethernet cable is
connected to the Power Supply Unit properly.
Or, make sure that the DI Unit AC adapter is
connected correctly.
Make sure that the DR Detector is connected
to the I/F cable correctly.
Make sure that the DR Detector is connected
to the I/F cable correctly. Or, make sure that
the Ethernet cable is connected correctly.
Make sure that the Ethernet cable for relying
serial exposure is connected to the Sync port.
Make sure the power cable is connected cor-
rectly.
Contact Konica Minolta technical
representatives.
Make sure that the Ethernet cable is
connected correctly.
Make sure that the Ethernet cable is
connected correctly. Or, make sure the power
switch of the AeroDR Interface Unit is on.
Contact Konica Minolta technical
representatives.
112
Corrective actions
Corrective actions
Corrective actions