Managing Results

Results are transmitted to the printer and computer as soon as all reagent
areas on the strip are read. If a record is flagged for a confirmatory report and
Edit flagged results is On, that record is not transmitted until after the end-of-
run reports complete.

End-of-Run Reports

The analyzer may display up to 3 end-of-run reports when the run, or a STAT
test, is completed. These reports display if you have marked any analytes to
flag for confirmatory or microscopic tests, and if Mark positives is On.
To request these reports:
1. Specify 1 or more tests for the Confirmatory Reports A and B or
Microscopy Report.
2. In the Setup routine, select
The Confirmatory and Microscopic Report screens display the
of the record, and the abbreviation for each positive analyte marked for
ID
flagging.
Up to 5 records may be displayed on 1 screen.
3. Use
and
Move Up
If both the Confirmatory and Microscopic Reports contain records, the
Confirmatory Reports display first.
4. Edit these results before exiting the Confirmatory Report. Refer to Editing
Results in the Confirmatory Reports‚ page 25.
5. Select
to print a report.
Print
6. Select
Return to Ready/Run
If an error is reported for 1 or more analytes, a report displays after the
Confirmatory and Microscopic Reports. This report displays last.
7. Retest any specimens listed.
Editing Results in the Confirmatory Reports
Use this procedure to edit the results of confirmatory testing.
1. During the end of run review, access the Confirmatory Report screens.
2. Select a record from the Confirmatory Report A screen.
The flagged positive test results display.
Clinitek Advantus Operator's Guide
for Edit flagged results.
On
to view additional records.
Move Down
to exit the report screen.