Setup Menu 5 - Siemens CLINITEK Advantus Operator's Manual

Hide thumbs Also See for CLINITEK Advantus:
Table of Contents

Advertisement

3. If necessary, select
4. When finished, select
Normal Range for SG and pH
You can select this option only if Mark positives is On.
Use this procedure to set the lower and upper limits of the normal range for SG
and pH. Set each limit separately. The upper limit must be higher than or equal
to the lower limit.
1. Select
Normal range for SG/pH
2. Select the
or
+
-
The limit changes by 1 reporting level until it is equal to the opposite limit or
is at the highest or lowest reporting level.
3. Select
Previous Screen
Normal Range for CRE
You can select this option only if you are using MULTISTIX PRO Reagent
Strips and Mark positives is On.
Use this procedure to select the lower and upper limits of the normal range for
creatinine. The upper limit must be equal to or higher than the lower limit.
1. Select
Normal range for CRE
2. Select the
or
+
-
The number changes by 1 reporting level until it is equal to the opposite
limit or is at the highest or lowest reporting level.
3. Select
Previous Screen

Setup Menu 5

At the fourth Setup menu, select
Menu Option
Color
Color choices
Clarity choices
Use default COL/CLA during run
Clinitek Advantus Operator's Guide
to display an additional screen of tests.
Next Screen
Previous Screen
.
keys next to each limit to raise or lower the limit.
.
.
keys next to each limit to raise or lower the limit.
.
Next Screen
.
to access the fifth Setup menu.
Default
Determined by analyzer
Yellow, Orange, Red, Green, Blue,
Brown, Other
Clear, SL Cloudy, Cloudy, Turbid,
Other
On
95

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents