Setup Menu 7 - Siemens CLINITEK Advantus Operator's Manual

Hide thumbs Also See for CLINITEK Advantus:
Table of Contents

Advertisement

3. For Prompted regular QC test or Compulsory regular QC test, enter the
interval:
a. Select
Set QC interval
b. Use the cycle key to select
c. Enter the QC interval.
d. Select
.
Enter

Setup Menu 7

At the sixth Setup menu, select
menu.
Menu Option
Microscopics setup
Edit flagged results
Enter sample IDs
Tech ID
Microscopics Setup
Enable entry of microscopics results.
1. Select
Microscopics setup
The display shows where to select 5 headings and their associated units.
The sixth selection enables you to enter custom data and units.
100
.
or
.
Hours
Days
to access the seventh Setup
Next Screen
Default
N/A
Off
Off
Off
.
Clinitek Advantus Operator's Guide