Completing Setup - Siemens CLINITEK Advantus Operator's Manual

Hide thumbs Also See for CLINITEK Advantus:
Table of Contents

Advertisement

Display
Use this test to ensure that all the lighted elements on the display are lit and
turned off properly.
1. Select
.
Display
The entire screen is lit for several seconds, then becomes blank.
This series is repeated twice more before returning to the previous menu.
2. If numerous faulty pixels exist, or irf they are located in critical areas, call
your local technical support provider to replace the display.
Printer
Use this test to ensure that the internal or 80-column external printer prints all
characters correctly.
1. If you are using an external printer, ensure that it is turned on.
2. Select
Printer.
3. Follow the directions on the screen.
4. Examine the printout for its readability.
The display automatically returns to the previous menu.

Completing Setup

1. When you have finished selecting the setup parameters, select
.
Ready/Run
The setup parameters are stored in the Clinitek Advantus analyzer.
Your changes are only saved when you select
NOTE:
.
Run
2. Print a copy of the setup report to verify your selections and retain in your
files:
a. At the Ready/Run screen, select
b. Select
.
Print
c. Select
Setup report
If printing from the internal printer, make a photocopy of the report,
NOTE:
because the thermal print may fade over time. If you have a new analyzer,
you can use the printout to select the parameters on the new analyzer.
110
.