Chapter 3
Theory of Operation

Scanning Modes

Single-Module Parallel Scanning

Four Isolated
Floating
Single-Ended
Inputs

Multiplexed Scanning

SCXI-1121 User Manual
There are four basic types of scanning modes possible with the
SCXI-1121—single-module parallel scanning, single-module multiplexed
scanning, multiple-module multiplexed scanning, and multiple-chassis
scanning (possible only with the SCXI-1001 chassis). For additional
information, consult Chapter 2,
Programming, your data acquisition board manual, and your SCXI chassis
user manual. If you need further information, contact National Instruments.
Single-Module parallel scanning is the simplest scanning mode. Directly
cable the SCXI-1121 to the data acquisition board as shown in Figure 3-7.
In this configuration, each analog signal has its own channel. Timing
signals are not necessary for this type of scanning because the module
provides all channels to the data acquisition board at all times. You can
implement single-module parallel scanning with any data acquisition board
that is appropriately cabled to the SCXI-1121.
SCXI-1121
CH0
MCH0
CH1
MCH1
CH2
MCH2
CH3
MCH3
Figure 3-7. Single-Module Parallel Scanning
Only the MIO-16 data acquisition boards support multiplexed scanning on
the SCXI-1121. During multiplexed scanning, a module sends the
SCANCLK signal to Slot 0 over the TRIG0 backplane line, and Slot 0
sends unique SCANCON signals to each module. Each module uses its
signal to reload MUXCOUNTER and to determine when the SCXI-1121
output is enabled. Slot 0 contains a module list first-in-first-out (FIFO)
memory chip, similar to the Channel/Gain FIFO on an MIO-16 board,
except that instead of having a channel number and gain setting for each
entry, the Slot 0 FIFO contains a slot number and a sample count for each
Configuration and
Cable Assembly
3-14
Installation, Chapter 5,
Data Acquisition Board
Analog Input 0
Analog Input 1
Analog Input 2
Analog Input 3
ni.com